Trilogy 100 Ventilator

Ventilator

Versatile, easy-to-use Trilogy 100 is designed for home, hospital and acute- care use to provide invasive and non-invasive ventilator support for adult and pediatric patients. A pressure and volume mode ventilator with proprietary algorithms, Trilogy 100 meets the changing needs of patients by combining unique flexibility with ease of use and advanced technology.

 

Specifications

Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Controls
Circuit types
  • Active with PAP; Passive
Ventilation types
  • Pressure and Volume
Volume modes
  • AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
  • CPAP; S; S/T; T; PC; PC-SIMV (w/PS); PC-MPV; AVAPS-AE
IPAP
  • 4 to 50 cm H?O
Pressure
  • 4 to 50 cm H?O
PEEP/EPAP for active circuits
  • 0 to 25 cm H?O
PEEP/EPAP for passive circuits
  • 4 to 25 cm H?O
CPAP for passive circuits
  • 4 to 20 cm H?O
Pressure support differential
  • 0 to 40 cm H?O
Alarms
Circuit disconnect
  • Off; 5 to 60 sec
Apnea rate
  • Off; 10 to 60 sec
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 40 to 2000 ml
High minute ventilation
  • Off; 1.0 to 99 l/min
Low minute ventilations
  • Off; 0.2 to 99 l/min
High respiratory rate
  • Off; 4 to 80 BPM
Low respiratory rate
  • Off; 4 to 80 BPM
Measured patient parameters
Respiratory rate
  • 0 to 80 BPM
Minute ventilation
  • 0 to 99 l/min
Leak rate
  • 0 to 200 l/min
Vte/Vti
  • 0 to 2000 ml
Peak inspiratory flow
  • 0 to 200 l/min
Peak inspiratory pressure
  • 0 to 99 cm H?O
Mean airway pressure
  • 0 to 99 cm H?O
% Patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Controls
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Ramp start pressure
  • 0 to 25 cm H?O for active circuits; 4 to 25 cm H?O for passive circuits; 4 to 19 cm H?O in CPAP
Ramp length
  • Off; 5 to 45 min
Flex
  • Off; 1 to 3
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 BPM