Trilogy EV300 Ventilator exclude humidifier
Trilogy EV300
Hospital ventilator
Trilogy EV300 has proven performance in both invasive and noninvasive ventilation
Transition to the future today with the next generation of Philips ventilators. The Trilogy EV300 ventilator delivers enhanced performance in noninvasive (NIV) and invasive (IV) ventilation, so patients can be treated with a single device throughout their hospital stay, regardless of changing conditions. Trilogy EV300 is designed to stay with your patients, saving your staff time and effort as patients move between hospital departments.
Features
Simplicity
- **Nominal run time per method in International Electrotechnical Commission (7.5 hour/battery). Detachable battery charge time 0% to 80% is 2.5 hours, internal battery charge time 0% to 100% is 3.5 hours.
Specifications
-
Ventilation modes A/C-PC -
Assist control (pressure control)
A/C-VC -
Assist control (volume control)
CPAP -
Continuous positive airway pressure
PSV -
Pressure support ventilation
S/T -
Spontaneous/timed ventilation
SIMV-PC -
Synchronized intermittent mandatory ventilation (pressure control)
SIMV-VC -
Synchronized intermittent mandatory ventilation (volume control)
-
-
Physical Weight -
6.3 Kg (13.8 lbs) with hot-swappable battery
Size -
19.3 cm D x 28.6 cm W x 24.5 cm H (7.6” D x 11.25” W x 9.65” H)
Screen dimensions -
20.32 cm (8")
Ingress protection -
IP22: protection against finger-sized objects & dripping water when tilted up to 15 degrees.
-
-
Oxygen Low flow -
0 to 30 l/min; maximum 10 psi
High pressure -
280 to 600 kPa (41 to 87 psi)
-
-
Magnet 1.5T I:E ratio -
9.9:1 to 1:9.9
-
-
Measured and displayed patient parameters Tidal volume (Vti or Vte) -
0 to 2000 ml
Minute ventilation (MinVent) -
0 to 30 l/min
Leak -
0 to 200 l/min
Respiratory rate (RR) -
0 to 90 BPM
Peak inspiratory flow PIF -
0 to 200 l/min
Peak inspiratory pressure PIP -
0 to 90 cmH?O
Mean airway pressure -
0 to 90 cmH?O
Spontaneous triggered breaths -
0 to 100 %
Dynamic compliance (Dyn C) -
1 to 100 ml/cmH?O
Dynamic resistance (Dyn R) -
5 to 200 cmH?O/l/sec
Dynamic plateau pressure -
0 to 90 cmH?O
Auto-PEEP -
0 to 20 cmH?O
FiO? with FiO? sensor -
21 to 100 %
SpO? with pulse oximeter -
0 to 100 %
Pulse rate with pulse oximeter -
18 to 321 bpm
EtCO? with CO? accessory -
0 to 150 mmHg
-
-
Electrical AC input voltage -
100V - 240V, 50/60 Hz, 1.7 - 0.6A
DC input voltage -
12/24V 6.5A
Li-ion batteries -
Internal/hot-swappable Li-ion batteries: 7.5 hours each nominal total run time (per IEC 80601-2-72)
Charge time for battery -
Charge time for detachable & internal battery: from 0% to 80%: 2.5 hours from 0% to 100%: 3.5 hours
-
-
Alarms Inspiratory Pressure -
1 - 90 cmH?O
Tidal Volume -
OFF, 10 - 2000 ml
Minute Ventilation -
OFF, 0.2 - 30 L/min
Respiratory Rate -
OFF, 1 - 90 BPM
Circuit Disconnection -
OFF, 5 - 60 sec
Apnea Interval -
5 - 60 sec
-
-
EasyDiagnost Eleva Storage temperature -
Temperature: -25°C to 70°C Relative humidity: 5% to 93% RH, non-condensing
-
-
Environmental Operating Conditions -
0°C to 40°C, 5%-90% RH (non-condensing), Atm.Pressure, 62 to 106 kPa, Altitude: -1261 to 12,971 ft
Transient operating temp -
-20°C to 50°C
-
-
Controls AVAPS with passive circuit -
PSV, S/T, and A/C-PC modes only
Tidal volume -
35 - 2000 ml on Dual Limb and Active Flow circuits, 50 - 2000 ml on passive and active PAP circuits
Breath rate -
0 - 80 BPM
PEEP -
0 - 35 cmH?0 for active circuits 3 - 25 cmH?0 for passive circuits
EPAP/CPAP -
3 - 25 cmH?0
IPAP -
3 - 60 cmH?0
Pressure support/control -
0 - 60 cmH?0
Inspiratory time -
0.3 - 5.0 sec
Rise time -
0-6 sec
Triggering and cycling -
Off, Auto-Trak, Sensitive Auto-Trak, and Flow Trigger
Flow trigger sensitivity -
0.5 - 9 l/min
Flow cycle sensitivity -
10% - 90% of peak flow
Flow pattern -
Square, Ramp
FiO? -
21% - 100%
Inspiratory time min/max -
0.3 - 3.0 sec
Backup ventilation -
ON - OFF
-
-
Standards General -
IEC 60601-1-1 Medical electrical equipment. Part 1-1: General requirements for safety
Collateral -
IEC 60601-1-11 Home Health Care Environment according to transit-operable usage
Particular -
ISO 80601-2-72 Part 2-72 ISO 80601-2-12 Part 2-12 ISO 80601-2-61 Part 2-61 ISO 80601-2-55 Part 2-55
Wireless communication -
Bluetooth v4.1, ISO/IEC 18092:2013, 21481 ed 2.0, 14443 ed 2.0, WLAN Std: IEEE 802.11 (2012) b/g/n
-